Chapter 7: Troubleshooting

Product Service Domestic 1-800-522-7658, International 816-333-8811


The following test allows you to check the pump and the connections
from the pump to the vacuum tube.

1. Place a rubber stopper in the vacuum port in the left side of the

collector chamber and another in the drain port in the bottom.

2. Restart the vacuum system and check your vacuum indication.

Insufficient vacuum indicates a bad pump or a leak in the
connections from the pump to the side of the chamber.





















The following test is used to check the collection chamber integrity.
The figure below points out potential areas to locate a leak:













Collector
Coil

Stopper

Areas for

Potential Leaks

Collector
Chamber