Labconco FreeZone and FreeZone Plus 6, 12 and 18 Liter Freeze Dry Systems 79600 Series User Manual

Page 56

Advertising
background image

52

Chapter 8: Modifying Your Freeze Dryer Display

Product Service: Domestic 1-800-522-7658, International 816-333- 8811


High line voltage to 0 V offset
Moisture sensor to enabled
Refrigeration to operate in Manual Mode


The display will default to the main freeze dry display information.
If you do not want to restore all factory settings, but want to adjust
offsets, press MENU briefly and the display will show:



Press SELECT until the appropriate offset is shown. Each
number represents 1°C

Press MENU and, if the drying chamber has been installed, the
display will show:



Press SELECT until the appropriate offset is shown. Each number
represents 1°C.

Press MENU and the display will show:





Press SELECT until the appropriate offset is shown. Each
number represents 0.001 mBar

Press MENU and the display will show





Press SELECT until the appropriate offset is shown. If the
Freeze Dryer is operating on nominal 115V, each number
represents 2V. If 3 is entered into the display, the low line
alarm will alarm when the voltage drops to approximately
97V, for example. 103-(2x3)=97V. If the Freeze Dryer is
operating on nominal 230V each number represents 4V (if
3 is entered into the display, the low line alarm will alarm
when the voltage drops to approximately 194V, for
example. 206-(4x3)=194V.

There will be no alarm indicated if DISABLED is selected.

COLLECTOR TEMP OFFSET:
O C

VACUUM OFFSET:
O uBar

LOW LINE OFFSET:
0

SDC TEMP. OFFSET:
O C