Calibration of Blood Pressure
To ensure correct measurements of simulated BP,
the sphygmomanometer used with the cuff needs to be
calibrated to the pressure sensor in the SimNewB system.
This is done by selecting the Calibrate BP function, inflating
the cuff to 100 mmHg and pressing the Calibrate quick
execution key as the pressure is held at exactly 100 mmHg.

Calibrating the Touchscreen Patient Monitor
- Double click on the Elo logo in the system tray in the lower right corner of the screen
- In the menu that appears select <Elo Touchscreen Properties>.
- Click the <Align> button in the dialog box.
- If target indicators first appear on the Instructor PC (laptop computer) screen, press Esc , or wait until
they appear on the Patient Monitor.
- When the target indicator appears on the Patient Monitor, touch each target as it changes position.
- Follow the on screen instructions then touch the green “checkbox” button.
- Click OK.

Connect the USB Camera to the USB HUB

Connect the Link Box to PC
- Connect the Network cable from the Link Box to the PC.


- Connect the audio cable (white mini jack with single wire) to the Link Box.
- Connect power supply or battery to the Link Box.

Connect the Manikin to the Link Box
- Connect the 37-Pin Serial Cable from the manikin to the Cable Adapter then connect the
Cable Adapter to the Link Box.
- Place the Blood Pressure cuff on the left arm of the manikin.
- Connect the clear tubing from the Blood Pressure cuff to the Link Box.
- Power on the Link Box.

Note:

-