Trouble-shooting – Laerdal SimNewB Advanced User Manual

Page 25

Advertising
background image

23

SimNewB™ manikin not breathing?
#1 Check that the chest rise is turned on in software.
#2 Check that the compressor is on and is connected to the

manikin.

#3 Check that there is a blood pressure - BP of 0/0 will set

the breathing rate to 0.

#4 Check that a perfusing rhythm is set.
#5 Check the batteries on the Link Box.

#6 Check that exhaust hose is not occluded or kinked.
#7 Check that lung closures are set to open or partial.
#8 Still not working, call your local Laerdal Technical Support

Office.

#1 Check the batteries for both units.

Compressor not working?
#1 Check setup instructions.
#2 Check the power connection to the compressor unit.
#3 Check that the source of electrical power works with

another device.

#4 Still not working: call your local Laerdal Technical Support

Office.

Manikin not making vocal sounds?
#1 Check that a breathing rate is set.
#2 Check that a BP is set.
#3 Check that a perfusing rhythm is set.
#4 Check that vocal sound volume is not set to 0.

Trouble-shooting