Oxygen Saturation

A standard SPO

2

probe (not included) can be attached to the

SimNewB system during the simulation to add visual realism
only. There is no interface between the SPO

2

probe and the

SimNewB system. It is up to the instructor to set the SPO

2

value that shall be visible to the simulation team members on
the optional monitor:

5. Circulation

Heart rate and Pulse
The SimNewB manikin has a heartbeat and palpatable umbilical
and right brachial pulses. The palpatable pulses can be switched
on or off by the instructor. The heartbeat can be listened to
using a stethoscope at the left side of the chest. The heart rate
can be varied between 0 and 300 bpm.

Even if the SimPad sets the pulse to off, palpating the

brachial pulse will turn on all pulses unless the BP cuff is inflated
to 10 mmHg or more.

NOTE:

Only use appropriate size neonate/infant stethoscope

I