Trouble-shooting – Laerdal SimNewB Advanced User Manual

Page 26

Advertising
background image

24

#1 Check that heart sound is selected.
#2 Check that heart rate is not set to 0.
#3 Check that heart sound volume is not set to 0.
#4 Check that perfusing heart rhythm is set.

#1 Check that lung sound is selected.
#2 Check that lung sound volume is not set to 0.
#3 Check that breathing rate is not set to 0.
#4 Check that lung closures are open or set to partial.

#1 Check that compressor is on and is connected to the

manikin.

#2 Check that pulse is enabled either through software or

brachial switch.

#3 Check that BP cuff pressure is below 10mmHg.

#4 Check that perfusing heart rhythm is not set to 0.

#5 Check that umbilical cord is inse rted all the way into

reservoir.

#1 Check that compressor is on and is connected to the

manikin.

#2 Check that motion is enabled in software.
#3 Check that exhaust hose is not occluded or kinked.
#4 Check that limbs are not bound by cables or blanket.

#5 Powder limb joint areas to reduce friction.

#1 Call your local technical product manufacturer for support.

Trouble-shooting

T

ROUBLE

-

S

HOOTING

C

OMPRESSOR

S

PECIFICATIONS

C

LINICAL

S

IMULATION

I

NTRODUCTION

S

ETUP

G

UIDE

Manikin not making heart sounds?

Manikin not making Lung sounds?

Manikin not having pulse?

Manikin not having good motions?

Laptop or any other peripherals not working?