Operation

External Calibration
The weight required for calibration/adjustment is defined in the factory settings (see “Specifications”).

Step Key (or instruction)

Display/Data output

1. Zero the balance

TARE

.

g


2. Start calibration

[CAL]

PLEASE WAIT

The balance will show
“PLEASE WAIT”,
after stable it will Display the
prompted calibration weight
(In this sample, 200g)

.

g CAL

4. Apply the prompted calibration

.

g

CAL

weight

5.Calibrate Weight

[CAL]

.

g

The balance will show the weighing
value and blink. After stable, the
balance will beep a sound and return
to the normal weighing mode to finish
the calibration

6. Remove the calibration weight.

.

g

11