Application programs – LW Measurements HRB 224 Magnetic Force Balance User Manual

Page 17

Advertising
background image

Application Programs

Weighing in Percent: ( Menu code: . . )
Purpose
This application program allows you to obtain weight readouts in percent which are in
proportion to a reference weight.

Features

í Storage parameter (rounding-off factor) for storing the reference weight

to calculate the percentage can be configured.

í Changing the reference sample percentage in the menu: reference the chapter Balance

setting.

í Optional automatic output of the reference weight to the data interface port when print

application parameters is set.

í Press and hold the (F) key (2 seconds) to set the reference percentage

í Toggling between percentage and weight by pressing (F).

Default/Initial parameter

Reference percentage 10( code: . . )

Preparation

Set parameters for the Weighing in percent program: reference Balance setting

Select the application program : Menu code: . . weighing in percent

Set the following parameters:

Code . .

5%

Code . .

10%

Code . .

20%

Code . .

50%

Code . .

100%

Please reference Balance setting

15