Troubleshooting guide – LW Measurements HRB 224 Magnetic Force Balance User Manual

Page 25

Advertising
background image

Troubleshooting Guide

Troubleshooting Guide
Error codes are shown on the main display for 2 seconds. The program then returns automatically to the
weighing mode

Display

Cause

Solution

No
segments appear on the
display

No AC power is available

The power supply is not
plugged in

Check the AC power supply

Plug in the power supply

The load exceeds the balance
capacity

Unload the balance

Something is touching the
weighing pan

Move the object that is
touching the weighing pan

Calibrate weight is wrong

Load on poise to recalibrate

Calibrate weight is unstable

Eliminate the unstable factor,
then recalibrate

poise weight not enough

Load on enough poise to
weighing

The maximum capacity
less than the data list in the
Technical specification

Switch on the balance without

the weighing pan on

Place the weighing pan and then

restart the balance with press

The weight readout
Is obviously wrong

The balance was not calibrated/
adjusted before weighing
The balance was not zeroed
before weighing

Calibrate/adjust the balance

Zero the balance before weighing