ASSEMBLE THE REACTOR
(
Continued
)
 
8. 
Connect cooling water to the magnetic 
drive. See Instruction Manual No. 
234M. 
 
9. 
Connect tubing to the rupture disc 
outlet and run to a safely vented area. 
See Instruction Manual 231M. 
 
10. Note the voltage and amperage 
requirement stamped on the controller 
data plate, and then plug the power 
cord into an appropriate outlet. Power 
for these reactors should be drawn 
from a grounded outlet capable of 
carrying up to the full current rating of 
the reactor. 
 
11. If an electric stirrer motor is supplied, 
turn the speed control knob fully 
counterclockwise on the Reactor 
Controller, turn on the motor switch and 
slightly increase the speed for a short 
run to check the stirrer drive system but 
do not turn on the heater, put heater 
toggle switch in center position (OFF). 
There must always be a vessel in the 
heater when it is turned on, and the 
vessel and heater sizes must match. If 
the heater is operated without proper 
size vessel in contact with the mantle, 
the mantle may overheat and fail. 
HEATERS
The 1 gallon and 1.5 gallon reactors are
equipped with 230 volt, 2800 watt heaters.
These heaters are designed to provide
uniform heat distribution to the vessels.
A thermowell is included in the heater to
accommodate an external thermocouple for 
contact with the outside vessel wall. 
Vessel Heaters Parts List
 
For Removable Vessels 
 
A1450HC3EE Heater 
Assembly, 1.0 G,
2800W,
230V
 
A1455HC3EE Heater 
Assembly, 1.5 G,
2800W,
230V
 
For Fixed Head Vessels 
 
A2870HC10EE Heater Assembly, 1.0 G, 
 2800W, 230V 
A2870HC20EE Heater Assembly, 1.5 G, 
 2800W, 
230V