PREFACE
Scope
These instructions describe the
installation, operation and maintenance of 
Parr Series 4580 HP/HT Pressure Reactors 
offered in two sizes, 1.0 and 1.5 gallon. They 
cover the basic steps to be followed when 
installing these reactors and describe the 
function of all standard components. They are 
intended to be used in conjunction with 
several related instruction sheets listed on the 
previous page. This information describes 
several components which are common to 
most Parr pressure reaction equipment, and 
includes safety precautions and other related 
information applicable to all research 
laboratories. The users should study all of 
these instructions carefully before starting to 
use these vessels so that they will fully 
understand the capabilities and limitations of 
the equipment. 
Safety Information
To avoid electrical shock, always: 
1. Use a properly grounded electrical outlet of 
correct voltage and current handling 
capability. 
2. Ensure that the equipment is connected to
electrical service according to local 
national electrical codes. Failure to 
properly connect may create a fire or 
shock hazard. 
3. For continued protection against possible
hazard, replace fuses with same type and 
rating of fuse. 
4. Disconnect from the power supply before
maintenance or servicing.
To avoid personal injury: 
1. Do not use in the presence of flammable 
or combustible materials; fire or explosion 
may result. This device contains 
components which may ignite such 
material. 
2. Refer servicing to qualified personnel.
General Specifications
Electrical Ratings
Controller ratings are found in the
Operating Instructions for the controller 
supplied with your reactor. 
Before connecting a controller to an
electrical outlet, the user must be certain that 
the electrical outlet has an earth ground 
connection and that the line, load and other 
characteristics of the installation do not 
exceed the following limits: 
 
Voltage: Fluctuations in the line voltage 
should not exceed 10% of the rated nominal 
voltage shown on the data plate. 
 
Frequency: Controllers can be operated from 
either a 50 or 60 Hertz power supply without 
affecting their operation or calibration. 
 
Current: The total current drawn should not 
exceed the rating shown on the data plate on 
the controller by more than 10 percent. 
 
Thermocouple: Unless otherwise specified, 
all Parr Controllers operate with a Type J 
(iron-constantan) thermocouple. The total 
resistance of the thermocouple and the lead 
wires should not exceed 20 ohms. If the 
resistance of the thermocouple circuit is 
higher, it will reduce the sensitivity of the 
control system.