Fluke 789 User Manual

Page 15

Advertising
background image

ProcessMeter™

Getting Acquainted with the Meter

7

Table 2. Input/Output Jacks

Item Jack Measurement

Functions

Source Current

Function

Simulate Transmitter

Function

A c

Input for current to 440 mA

continuous. (1 A for up to 30

seconds.) Fused with a 440 mA

fuse.

Output for dc current to 24 mA.

Output for loop power supply.

mA d

Input for current to 30 mA.

Fused with a 440 mA fuse.

Common for dc current output to

24 mA. Common for loop power

supply.

Output for transmitter

simulation to 24 mA. (Use in

series with an external loop

supply.)

V

Input for voltage to 1000 V,

Ω

,

continuity, and diode test.

COM

Common for all measurements.

Common for transmitter

simulation to 24 mA. (Use in

series with an external loop

supply.)