Loop power supply mode – Fluke 789 User Manual

Page 37

Advertising
background image

ProcessMeter™

Loop Power Supply Mode

29

Loop Power Supply Mode

The Loop Power Supply Mode can be used for powering
up a process instrument (transmitter). While in Loop
Power Mode, the meter acts like a battery. The process
instrument regulates the current. At the same time, the
meter measures the current that the process instrument is
drawing.

The meter supplies loop power at a nominal 24 V dc. An
internal series resistance of 250

Ω

can be switched in for

communication with HART and other smart devices by
pressing J (Blue). Pressing J (Blue) again switches out
this internal resistance.

When loop power is enabled, the meter is configured to
measure mA and > 24 V dc is sourced between the mA
and A jacks. The mA jack is the common and the A jack is
at > 24 V dc. Connect the meter in series with the
instrument current loop as Figure 10 shows.

32

30

28

26

24

22

20

0

4

8

12

16

20

24

Loop voltage w/ 250

Loop voltage w/o 250

Voltage (V)

Current (mA)

anw020f.eps

Figure 9. Loop Power Voltage vs. Current