Specifications – Fluke 789 User Manual

Page 47

Advertising
background image

ProcessMeter™

Specifications

39

Specifications

All specifications apply from +18

°

C to +28

°

C unless

stated otherwise.

All specifications assume a 5-minute warm-up period.

The standard specification interval is 1 year.

Note

“Counts” refers to the number of increments or
decrements of the least significant digit.

DC Volts Measurement

Range (V dc)

Resolution

Accuracy,

±

(% of Reading + Counts)

4.000

0.001 V

0.1 % + 1

40.00

0.01 V

0.1 % + 1

400.0

0.1 V

0.1 % + 1

1000

1 V

0.1 % + 1

Input impedance: 10 M

Ω

(nominal), < 100 pF

Normal mode rejection ratio: > 60 dB at 50 Hz or 60 Hz
Common mode rejection ratio: > 120 dB at dc, 50 Hz, or 60 Hz
Overvoltage protection: 1000 V