Frequency Counter Accuracy

Range Resolution

Accuracy,

±

(% of Reading + Counts)

199.99 Hz

0.01 Hz

0.005 % + 1

1999.9 Hz

0.1 Hz

0.005 % + 1

19.999 kHz

0.001 kHz

0.005 % + 1

Display updates 3 times/second at > 10 Hz

Frequency Counter Sensitivity

Input Range

Minimum Sensitivity (rms Sinewave)

5 Hz to 5 kHz*

AC

DC

(approximate trigger level 5 % of full scale)

400 mV

150 mV (50 Hz to 5 kHz)

150 mV

4 V

1 V

1 V

40 V

4 V

4 V

400 V

40 V

40 V

1000 V

400 V

400 V

*Usable 0.5 Hz to 20 kHz with reduced sensitivity.