Measuring electrical parameters, Input impedance, Ranges – Fluke 789 User Manual

Page 26: Testing diodes

Advertising
background image

789
Users Manual

18

Measuring Electrical Parameters

The proper sequence for taking measurements follows:

1. Plug the test leads into the appropriate jacks

2. Set the rotary function switch to the desired function

3. Touch the probes to the test points

4. View the results on the LCD display

Input Impedance

For the voltage measurement functions, input impedance
is 10 M

Ω

. See "Specifications" for more information.

Ranges

A measurement range determines the highest value and
resolution at which the meter can measure. Most meter
measurement functions have more than one range (see
"Specifications").

Make sure the correct range is selected:

If the range is too low, the display shows OL
(overload).

If the range is too high, the meter will not be
displaying its most precise measurement.

The meter normally selects the lowest range that will
measure the applied input signal (Auto Range showing on
the display). Press R to lock the range. Each time
R is pressed, the meter selects the next higher range.
At the highest range, it returns to the lowest range.

If the range is locked, the meter resumes auto ranging
when it is changed to another measurement function or
when R is pressed and held for 1 second.

Testing Diodes

To test a single diode:

1. Insert the red test lead into the Vjack and

black test lead into the COM jack.

2. Set the rotary function switch to

V

.

3. Press

J (Blue) so that the

D symbol is on the

display.

4. Touch the red probe to the anode and the black

probe to the cathode (side with band or bands). The
meter should indicate the appropriate diode voltage
drop.

5. Reverse the probes. The meter displays OL,

indicating a high impedance.