ABUS Technologies DP Series Smart Pressure Transmitter User Manual

Page 8

Advertising
background image

ABUS TECHNOLOGIES INC.

M94

8

2.4.1 Push Button Functions

1. Zero & Span Shift

Zero Shift : push key S & Z at the same time (key S on the left while key Z on the right) for 6
seconds, the screen indicates “Hart”, means zero and span adjustment status is activated; push
key Z for 5 seconds, “Hart” on the screen disappeared; zero shifts successfully and exits the
status of activation.
Span Shift: push key S & Z at the same time (key S on the left while key Z on the right) for 6
seconds, the screen indicates “Hart”, means zero and span adjustment status is activated; after
confirming the pressure of equipment is span pressure, push key S for 6 seconds; then “Hart” on
the screen disappeared; span shifts successfully and exits the status of activation.
Note: After entering the status of activation, if you don’t want to adjust, push key S & Z together;
after loosing the key, “Hart” on the screen disappears; it means having exit the status of
activation

.

2. Parameter Setting

2-1 keys : Push key S & Z = exit key; exit key comes into force after loosing key S & Z; push key
Z for 4 seconds to shift
2-2 Parameters : After pushing key S for 6 seconds, the screen indicates “PASS”; then push key
Z, the screen indicates value and the cursor flickers. The flickering bit is modifier bit; push key Z
the value increases 1 but not carry. Push key Z for 4 seconds the cursor shifts. Follow the above
mentioned operation to revise the value to be 160, then push key S to enter adjustment procedure
of next parameter. Meanings of parameters are as follows

:

Parameter Prompt

Value

Meaning

PASS

XXXXX

Password Verification, PW: 160

bdSh

XXXX.X

Process value to 20 mA (full scale value)

bdSL

XXXX.X

Process Value to 4 mA (zero value)

trAn

XXXXX

Transmitting function, “=0” stands for linearity, “=1” stands for
evolution

UnIt

XXXXX

Indicating units, 0=KPa, 1=KPa, 2=KPa, 4=KPa, 6=Pi, 7=bar,
8=KPa, 9=KPa, 11=Pa, 12=KPa, 13=KPa, 15=MPa, 16=mA,
17=%, 18=S, 19=m

3

, 20=t/h

dAhP

XXXX.X

Damping time, max limited value is 30 s when program runs

PrEL

XXXXX

Low point trimming. Process value adjusted to input value; after
pushing key Z revise low point trimming pressure value; don’t
push key Z if you don’t want to adjust; directly push key S to
enter next parameter

PrEh

XXXXX

High point trimming. Process value adjusted to input value;
after pushing key Z revise high point trimming pressure value;
don’t push key Z if you don’t want to adjust; directly push key S
to enter next parameter

InIt

yes/no

Hart. Three-point linearization and trimming data initialization

Hotkey: push key Z for 6 seconds to do zero trimming
Note: Please operate carefully for italic part parameters; make sure correct value of pressure
forced on equipment

DETAILS

1 Rotating Display

2 Bright White Backlit Display

3 Bar Graph Display

4 Display in various Engineering Units

5 Zero Button

6 Span Button

4

2

1

3

5

6

Advertising