Page 21: Caution

Advertising
background image

KB (E3.1 + E5.1) 11/2014

page 21/100

If you need to return the unit, please use the original packing and observe the guidelines for safe lifting

and transportation (chap. 3.2).
For disposal of the transport packing, see chap. 18.1.
Note on second-hand units (Ex-Demo-Units):
Second-hand units are units that were used for a short time for tests or exhibitions. They are thoroughly

tested before resale. BINDER ensures that the chamber is technically sound and will work flawlessly.
Second-hand units are marked with a sticker on the unit door. Please remove the sticker before

commissioning the unit.

3.2 Guidelines for safe lifting and transportation

The front castors

of

units size 240, 400, and 720 can be blocked by brakes. Please move the units with

castors only when empty and on an even surface, otherwise the castors may be damaged. After

operation, please observe the guidelines for temporarily decommissioning the unit (chap. 18.2).

CAUTION

Sliding or tilting of the unit.
Damage to the unit.
Risk of injury by lifting heavy loads.

Transport the unit in its original packaging only.

For moving or shipping, secure the unit with transport straps.

Do NOT lift or transport the unit using the door, the handle, or the lower housing.

Do NOT lift units size 720 by hand.

Lift units size 23. 53 und 115 with the aid of four people in the area of all 4 unit feet.

Lift units size 240 with the aid of 6 people or with a fork lifter. Set the fork lifter only

from the front or rear in the middle of the unit.

Lift units size 720 using technical devices (fork lifter). Set the fork lifter only from the

front or rear in the middle of the unit.

Do NOT set the fork lifter from the unit side.

Permissible ambient temperature range during transport: -10 °C / 14°F to +60 °C / 140°F.

You can order transport packing for moving or shipping purposes from BINDER service.

3.3 Storage

Intermediate storage of the unit is possible in a closed and dry room. Observe the guidelines for

temporary decommissioning (chap. 18.2).

Permissible ambient temperature range during storage: -10 °C / 14°F to +60 °C / 140°F.

Permissible ambient humidity: max. 70 % r.H., non-condensing

When after storage in a cold location you transfer the unit to its warmer installation site, condensation may

form. Before start-up, wait at least one hour until the chamber has attained ambient temperature and is

completely dry.