BINDER KB 23 User Manual

Page 61

Advertising
background image

KB (E3.1 + E5.1) 11/2014

page 61/100

2. Entering the time program to the program editor

Program table for the example program:

Section


SEC

Temperature

set-point

[ °C]

TEMP

Section length

[hh.mm]

TIME

Fan speed

[%]

FAN

Operation lines


O.LINE

S01

20

119:59

100

000

S02

20

000:01

100

000

S03

5

047:59

100

000

S04

5

000:01

100

000

In this example the program will be entered to the first program place (P01).

Normal display

Press key

for 5 sec.

Display 1 shows

e.g. 0000

Display 2 shows PROGRAM EDITOR

(you are in the program editor)

Press program key

Display 1 shows

0000

Display 2 shows

UserCod? 0000

(enter user code)

Enter user code using arrow keys

e.g.

0001

(basic setting, adjustable in the user level,

chap. 10).

Value is shown in both displays.

Automatically forward after 2 sec.

Display 1 shows

01

program P01 selected

Display 2 shows

alternating

---:--- PRG.

program can be selected

CONTINUE X/W

(information: to 1

st

program section with X/W)

Select program

P01 using arrow keys

Value is shown in Display 1.

Press key

In the selected program P01 the first program section S01 is shown:

Display 1 shows

01

section S01 has been selected

Display 2 shows

alternating

P01: --- SEC.

enter new set-points for the individual variables with button

X/W.

CONTINUE X/W

or

NEW SEC. X/W

Select section S01using arrow keys

Press key

Display 1 shows

20.0

C

(actual temperature set-point)

Display 2 shows

alternating

S01: TEMP 20.0

(variable: temperature in °C)

CONTINUE X/W

(information: go on with X/W)

Entry of temperature set-point of

20 °C

using arrow keys