20.4 KB (E3.1) technical data

Unit size

23

53

115

Exterior dimensions
Width

mm / inch

433 / 17.05

634 / 24.96

834 / 32.83

Height (incl. feet)

mm / inch

618 / 24.33

837 / 32.95

1022 / 40.24

Depth

mm / inch

516 / 20.31

576 / 22.68

646 / 25.43

incl. door handle, I-panel, connection

mm / inch

73 / 2.87

100 / 3.94

100 / 3.94

Wall clearance

mm / inch

100 / 3.94

100 / 3.94

100 / 3.94

Wall clearance door open

mm / inch

100 / 3.94

160 / 6.30

160 / 6.30

Steam space volume

l / cu.ft.

36 / 1.27

77 / 2.72

158 / 5.58

Interior dimensions
Width

mm / inch

222 / 8.74

400 / 15.75

600 / 23.62

Height

mm / inch

330 / 12.99

400 / 15.75

480 / 18.90

Depth

mm / inch

277 / 10.91

330 / 12.99

400 / 15.75

Interior volume

l / cu.ft.

10 / 0.35

53 / 1.9

115 / 4.1

Number of racks standard / max.

2/3

2/4

2/5

Load per rack

Kg / lbs

12 / 26

15 / 33

20 / 44

Permitted total load

Kg / lbs

25 / 55

40 / 88

50 / 110

Weight (empty)

Kg / lbs

44 / 97

72 / 159

105 / 232

Number of door(s)

number

1

1

1

Inner glass door(s)

number

1

1

1

Temperature data

Temperature range 1)

°C

°F

0 to +100

4 to 212

-5 to +100

23 to 212

-5 to +100

23 to 212

Temperature fluctuation

at 5 °C

± K

0.2

0.1

0.1

at 40 °C

± K

0.2

0.1

0.1

Temperature uniformity

(variation)

at 5 °C

± K

0.7

0.6

0.4

at 25 °C

± K

0.3

0.2

0.1

at 40 °C

± K

0.3

0.3

0.2

Heating up time 2)

to 40 °C

min

11

5

9

Cooling down time from 40 °C to 5 °C

min

61

58

83

Recovery time after doors

were open for 30 sec. 2)

at 5 °C

min

5

4

5

at 40 °C

min

2

1

1

Electrical data KB
IP System of protection acc. to EN 60529

20

20

20

Nominal voltage (±10 %)

V

230 1N~

230 1N~

230 1N~

Power frequency

Hz

50/60

50

50

Power plug

shock proof plug

Nominal power

kW

0.34

0.46

0.46

Energy consumption 3)

at 5 °C

Wh/h

50

260

222

at 40 °C

Wh/h

60

215

115

Overvoltage category acc. to IEC 61010-1

II

II

II

Pollution degree acc. to IEC 61010-1

2

2

2

Unit fuse 5x20 mm 230V / middle-time-lag

M

Amp

10

10

10

external

external

external