10. User level

In this menu the following parameters can be se (in brackets the corresponding abbreviated information

given in display 2):
•

Unit address (Address)
Setting of controller address (1 to 254) for operation with the communication software APT-COM™.

•

User code (User-cod)
Modification of the user code setting (factory setting “0001”) for access to the user level and the

program editor.

Keep in mind any modification of the user code. Without a valid user code it is no longer

possible to access these levels.

•

Safety controller mode (Saf.mode)
The set-point mode can be entered as ...
Limit: Maximally permitted temperature as absolute value (example: temperature set-point 37 °C /

98.6°F, limit setting to 39 °C / 102.2°F)
Offset: Maximally permitted temperature exceeding of the actual set-point temperature (e.g. 2 °C).

This maximally permitted temperature automatically changes with every change of the temperature

set-point.

•

Safety controller set-point (Saf.setp)
The temperature set-point setting of the safety controller (over temperature safety device class 3.1) is

shown and can be changed. An absolute value (e.g. 40 °C / 104°F) in case of setting “limit” or a

relative value (e.g. 2 °C) in case of setting “Offset” can be entered.

Regularly check the setting of the set-point mode and of the over temperature safety

device set-point entered in this mode for the temperature value in Fixed value operation or

for the maximum temperature value of the selected temperature program in Program

operation mode.

•

Decimal point position (Decimal)
Selection if integer values or one position after the decimal point can be entered. The integer

representation is shown in Display 2 (set-point entry) while the actual value in Display 1 is always

displayed with one decimal point.

•

Audio Alert (Buzzer)
Inactive: no audible signal (buzzer) in case of an alarm event.
Active: in case of an alarm event (see chap. 13.2) an audible signal (buzzer) will sound. It can be reset

by pressing the “

EXIT” button.

•

Selection of controller menu language (Language)
German, English, or French can be selected.

•

Counter of operating hours (Oper.hs)
Information about the number of operating hours currently reached or since the last reset. (no setting,

display only).

•

Max. number of operating hours (Op.limit)
Entry of a limit number of operating hours, i.e., the maximum number of operating hours that can be

run. Maximum setting: 9999. Reaching the limit has no effect.

•

Reset operating hours (Op.back)
Reset operating hours to zero.