3 proceeding in detail – BINDER KB 23 User Manual

Page 59

Advertising
background image

KB (E3.1 + E5.1) 11/2014

page 59/100

3. Set the number of cycles to infinite in the program start level and start the program

The described example program must be started once at the precise moment of temperature

change (on Monday e.g., at 0.01 or at 7.00). If the program cannot be manually started at the

desired moment, you can program a suitable program delay-time of 99 hs. 59 min. max. After

rundown of this delay time, the program will start automatically (chap. 9).

12.3 Proceeding in detail

1. Settings in the user level:

Setting the safety controller to “Limit” and 3 °C above the maximum temperature value of the

program
The temperature set-point of the safety controller (over temperature protection class 3.1) is shown and

can be changed. An absolute value in case of setting “limit”, or a relative value in case of setting

“Offset” can be entered Select setting “Limit” and enter a value, which is by 3 °C above the maximum

value (i.e., 23 °C / 73.4°F).

Setting the maximum section duration (Prg. Time) to 999 hs. 59 min.
The maximum length of a program section can be set – in common for all program sections – to 99 hs

59 min or to 999 hs 59 min. Select setting 999:59.

When changing the maximum duration setting, pre-existing programs will be deleted in

the program editor.


Switching off the tolerance limits function
Select setting “0” meaning tolerance limits off. Thus an interruption of the program course during the

heating-up or cooling-down phases during the rapid “set-point step” phase is avoided.

Deactivating the week program timer
Deactivate the week program timer before entering a program (factory setting). Otherwise, any setting

of the operation lines in the program editor is ineffective.

Normal Display

Display 1 shows

e.g. 19.8

(actual temperature value)

Display 2 shows e.g. 15.05.06 13:52 - -

(actual date and time, actual switching state of week

program timer channel 1: Off, channel 2: Off)

Press down key

for 5 sec

Display 1 shows

e.g. 0000

Display 2 shows

PROGRAM EDITOR

(you are in the program editor)

Press down key

for 5 sec

Display 1 shows

0000

Menu visible only if week program timer is activated

Display 2 shows WEEK PROG. EDITOR

(you are in the week program editor)

Press down key

for 5 sec

Display 1 shows

0000

Display 2 shows

USER-LEVEL

(you are in the user level)

Press program key

Advertising