VOLT SURGE
This test will produce voltage surge defined by Table 8-16. This test will be repeated three times
with ten seconds intervals. The voltage values are halved for 14.0V category B equipment.
Category
Surge 1
Surge 2
Volt
Dwell(msec)
Volt
Dwell(msec)
A
46.3
100
37.8
1000
B
60
100
40
1000
Z
80
100
48
1000
Table 8-16: Abnormal Voltage Surge
8.1.1.6.3 Emergency Test
The Emergency test is selected from the DO-160 DC Main Menu. This test will set the output
voltage to 18V for 28V equipment and to 9V for 14V equipment. The test will last for 30 minutes.
The
←
key (backspace) will terminate the test at any time.