ARCHMI-9XX Series User Manual

97

This function offers two ways to calibrate your touch screen. ‘Standard Calibration’

adjusts most touch screens. ‘Advanced Calibration’ adjusts aging touch screens.

Standard Calibration Click this button and arrows appear pointing to red

squares. Use your finger or stylus to touch the red
squares in sequence. After the fifth red point calibration

is complete. To skip, press ‘ESC’.

Advanced Calibration Advanced Calibration uses 4, 9, 16 or 25 points to

effectively calibrate touch panel linearity of aged touch

screens. Click this button and touch the red squares in
sequence with a stylus. To skip, press ESC’.

Step 1. Please select a device then click “Configure”. You can also double click the
device too.

Step 2.Click “Standard Calibration” to start calibration procedure