Electromagnetic compatibility information – Beurer BC 58 User Manual

Page 54

Advertising
background image

54

Electromagnetic Compatibility Information

Table 1
For all ME EQUIPMENT and ME SYSTEMS

Guidance and manufacture’s declaration – electromagnetic emissions

The BC 58 is intended for use in the electromagnetic environment specifi ed below.
The customer or the user of the BC 58 should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment – guidance

RF emissions CISPR 11

Group 1

The BC 58 uses RF energy only for its internal function. Therefore, its RF emissi-
ons are very low and are not likely to cause any interference in nearby electronic
equipment.

RF emissions CISPR 11

Class B

The BC 58 is suitable for use in all establishments other than domestic and
those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.

Harmonic emissions IEC 61000-3-2

Not applicable

Voltage fl uctuations/fl icker emissions
IEC 61000-3-3

Not applicable

Table 2
For all ME EQUIPMENT and ME SYSTEMS

Guidance and manufacturer’s declaration – electromagnetic immunity

The BC 58 is intended for use in the electromagnetic environment specifi ed below. The customer or the user of the BC 58 should assure
that it is used in such an environment.

IMMUNITY test

IEC 60601
test level

Compliance
level

Electromagnetic environment – guidance

Electrostatic discharge (ESD)
IEC 61000-4-2

± 6 kV contact
± 8 kV air

± 6 kV contact
± 8 kV air

Floors should be wood, concrete or ceramic tile. If fl oors are covered
with synthetic material, the relative humidity should be at least 30 %.

Power frequency (50/60 Hz)
magnetic fi eld IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fi elds should be at levels characteristic of
a typical location in a typical commercial or hospital environment.

NOTE: U

T

is the a.c. mains voltage prior to application of the test level.