Table 3
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING

Guidance and manufacturer’s declaration – electromagnetic immunity

The BC 58 is intended for use in the electromagnetic environment specifi ed below. The customer or the user of the BC 58 should assure
that it is used in such an environment.
IMMUNITY
test

IEC 60601
test level

Compliance
level

Electromagnetic environment – guidance

Conducted RF
IEC 61000-4-6

Radiated RF
IEC 61000-4-3

3 Vrms
150 kHz to
80 MHz

3 V/m
80 MHz to
2.5 GHz

3 V

3 V/m

Portable and mobile RF communications equipment should be used no closer to any part
of the BC 58, including cables, than the recommended separation distance calculated from
the equation applicable to the frequency of the transmitter.

Recommended separation distance:

d = 1.2

d = 1.2

80 MHz to 800 MHz

d = 2.3

800 MHz to 2.5 GHz

Where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and d is the recommended separation distance in meters (m).

Field strengths from fi xed RF transmitters, as determined by an electromagnetic site survey,

a

should be less than the compliance level in each frequency range.

b

Interference may occur in the vicinity of equipment marked with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is aff ected by absorption and refl ection from

structures, objects and people.