Beurer BM 90 User Manual

Page 100

Advertising
background image

100

Table 3

For Me eQuIPMeNT and Me sYsTeMs that are not LIFe-suPPORTING

Guidance and manufacturer’s declaration - electromagnetic immunity

The BM90 is intended for use in the electromagnetic environment specified below. The customer or the user

should assure that it is used in such an environment.

IMMuNITY

test

IeC 60601 test level

Compliance

level

electromagnetic environment - guidance

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 V

rms

150 kHz to 80

MHz

3 V/m 80 MHz

to 2.5 GHz

3 V

rms

3 V/m

Portable and mobile RF communications equip-

ment should be used no closer to any part of the

BM90, including cables, than the recommended

separation distance calculated from the equation

applicable to the frequency of the transmitter.

Recommended separation distance

d = 1.167 P

d = 1.167 P 80 MHz to 800 MHz

d = 2.333 P 800 MHz to 2.5 GHz

where P is the maximum output power rating of

the transmitter in watts (W) according to the trans-

mitter manufacturer and d is the recommended

separation distance in metres (m).

Field strengths from fixed RF transmitters, as

determined by an electromagnetic site survey,

a

should be less than the compliance level in each

frequency range.

b

Interference may occur in the vicinity of equip-

ment marked with the following symbol:

NOTe 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTe 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorpti-

on and reflection from structures, objects and people.

Advertising