In such cases, repeat the measurement. Ensure that the cuff line is correctly attached and that you do not move

or speak. If necessary, reinsert or replace the batteries.

10. Cleaning and storing the unit

Clean your blood pressure monitor carefully using a slightly damp cloth only.

Do not use any cleaning agents or solvents.

Under no circumstances should you hold the unit under water, as this can cause liquid to enter and damage

the unit.

When you store the unit, no heavy objects should be placed on top of it. Remove the batteries. The cuff line

should not be bent sharply.

11. Technical specifications

Model no.

BM 90

Measurement method

Oscillometric, non-invasive blood pressure measurement on the upper

arm

Measurement range

Cuff pressure 0–300 mmHg, systolic 30–280 mmHg, diastolic 30–280

mmHg, pulse 40–199 beats per minute

Display accuracy

systolic ± 3 mmHg/diastolic ± 3 mmHg/

pulse ± 5% of the value shown

Measurement inaccuracy

max. permissible standard deviation according to clinical testing:

systolic 8 mmHg/diastolic 8 mmHg

Memory

60 memory spaces

Dimensions

L 140 mm x W 60 mm x H 56 mm

Weight

Approximately 471 g (without batteries)

Cuff size

22 to 42 cm

Permissible operating conditions

+5°C to +40°C, ≤ 80% relative humidity (non-condensing)

Permissible storage conditions

-20°C to +60°C, < 10–93% relative humidity
Ambient pressure: 800–1050 hPa

Power supply

6V

(4x 1.5V AAA)

Battery life

For approx. 120 measurements, depending on levels of blood pressure

and pump pressure

Radio frequency

RF 868MHz

Protection class

Internal supply, IPX0, no AP or APG, continuous operation

Classification

Application part type B