a

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land

mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically

with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic

site survey should be considered. If the measured field strength in the location in which the BM90 is used

exceeds the applicable RF compliance level above, the BM90 should be observed to verify normal operation.

If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating

the BM90.

b

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.