Table 2

For all Me eQuIPMeNT and Me sYsTeMs

Guidance and manufacturer’s declaration - electromagnetic immunity

The BM90 is intended for use in the electromagnetic environment specified below. The customer or the user of

BM90 should assure that it is used in such an environment.

IMMuNITY test

IeC 60601 test level

Compliance level

electromagnetic environ-

ment - guidance

Electrostatic

discharge (ESD)

IEC 61000-4-2

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

Floors should be wood, con-

crete or ceramic tile. If floor

are covered with synthetic

material, the relative humidity

should be at least 30%.

Electrical fast

transient/burst

IEC 61000-4-4

±2 kV for

power supply lines

±2kV for

power supply lines

Mains power quality should

be that of a typical commer-

cial or hospital environment.

Surge IEC

61000-4-5

±1 kV line(s)

to line(s)

±1 kV

differential mode

Mains power quality should

be that of a typical commer-

cial or hospital environment.

Voltage dips, short in-

terruptions and voltage

variations on power

supply input lines IEC

61000-4-11

<5% U

T

(>95% dip in U

T

)

for 0.5 cycle

<5% U

T

(>95% dip in U

T

)

for 0.5 cycle

Mains power quality should

be that of a typical commer-

cial or hospital environment.

If the user of the BM90

requires continued operation

during power mains inter-

ruptions, it is recommended

that the BM90 be powered

from an uninterruptible power

supply or a battery.

40% U

T

(60% dip in U

T

)

for 5 cycles

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

<5% U

T

(>95% dip in U

T

)

for 5 sec

<5% U

T

(>95% dip in U

T

)

for 5 sec

Power frequency

(50Hz) magnetic

field IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic

fields should be at levels

characteristic of a typical lo-

cation in a typical commercial

or hospital environment.

NOTe U

T

is the a.c. mains voltage prior to application of the test level.