6.2 Displaying average blood sugar levels

You can display the average blood sugar value from the last 7, 14, 30 and 90 days.

1

The measuring device must be switched off. Press the rocker button S [5].

The start display appears briefly.
Press the rocker button S [5] again.
The blood sugar level unit, "7 days" and the average value are displayed.

2

PressS [5] repeatedly to display the average value for 14, 30 and 90 days.

3

You can cancel the process at any time. To do so, press the ON/OFF button or wait
until the unit switches off automatically after 1 minute.

Item Meaning

1 Average

value

2

Number of days, e.g. 7

6.3 Deleting the measured value memory

1

The measuring device must be switched off.

2

Remove the batteries from the battery compartment.

3

Simultaneously press the S button [5] and the ON/OFF button [3] and hold them
down.

4

Reinsert the batteries while holding down the buttons; pay attention to the correct po-
larity.

5

After inserting the batteries, continue holding down the S button and the on/off button
[3] for another five seconds. "dEL" is displayed and the measured values are deleted.