Beurer GL 40 (mmoL) User Manual
Page 34
34
Beurer GL40 mmol/L
9 Technical
specifications
Dimensions (WHD)
47 x 85 x 14 mm
Weight
43 g (including batteries)
Electrical power
2 x 3V CR2032 button cell batteries
Battery life
More than 1000 measurements
Measured value memory
480 measured values with date/time
Data retention when replacing the batteries
Average values
Blood sugar: for 7, 14, 30, 90 days
Automatic switch-off
2 minutes after the last operation
Storage/
transport temperature
Temperature: +2°C – +30°C
Relative humidity: < 90%
Operating ranges
Temperature: +10°C – +30°C
Relative humidity: < 90% non-condensing
Glucose measurement
range
Glucose: 1.1 – 35 mmol/L
Blood sample
Capillary whole blood
Blood quantity
0.6 microlitres
Measuring time for blood
sugar
Approx. 5 seconds
Calibration
Plasma
Test method
Amperometric biosensor
Application
Suitable for private, individual use
System function test
Each time the unit is switched on