8 troubleshooting – Beurer GL 40 (mmoL) User Manual

Page 32

Advertising
background image

32

Beurer GL40 mmol/L

8 Troubleshooting

Messages on the display for batteries and blood sugar measurement
No.

Cause

Action

LP

Flat batteries

Replace all batteries.

Ht

Temperature of the measure-
ment room, measuring device
or test strip above the allow-
able limit.

Repeat test with new test strip as soon as the
measurement room, measuring device and test
strip have reached room temperature (+20°C to
+25°C).

Lt

Temperature of the measure-
ment room, measuring device
or test strip below the allow-
able limit.

Repeat test with new test strip as soon as the
measurement room, measuring device and test
strip have reached room temperature (+20°C to
+25°C).

Err

Used or contaminated
test strip was inserted.

• Insert test strip which has not been used or

spoiled.

• Repeat blood sugar measurement.

001 System

error

• Remove the batteries and reinsert them.

Contact our customer service department if
the problem persists.

005 Memory

error

• Remove the batteries and reinsert them.

Contact our customer service department if
the problem persists.