ELECTROMAGNETIC COMPATIBILITY INFORMATION

Guidance and manufacturer’s declaration – electromagnetic immunity

The “Mini Daylight” is intended for use in the electromagnetic environment specified below. The customer

or the user of the “Mini Daylight” should ensure that it is used in such an environment.

Immunity test

IEC 60601

test level

Compliance

level

Electromagnetic environment –

guidance

Electrostatic

discharge (ESD)

IEC 61000-4-2

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

Floors should be wood, concrete or

ceramic tile. If floors are covered with

synthetic material, the relative humidity

should be at least 30 %.

Electrical fast

transient/burst

IEC 61000-4-4

±2 kV for power

supply lines

±1 kV for input/output

lines

±2 kV for power

supply lines

±1 kV for input/

output lines

Mains power quality should be that of

a typical commercial or hospital envi-

ronment.

Surge

IEC 61000-4-5

±1 kV differential mode

±2 kV common mode

±1 kV differential

mode

±2 kV common

mode

Mains power quality should be that of

a typical commercial or hospital envi-

ronment.

Voltage dips,

short interruptions

and voltage vari-

ations on power

supply input lines

IEC 61000-4-11

<5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5 sec

<5 % U

T

(>95 % dip in U

T

)

for 0,5 cycle

40 % U

T

(60 % dip in U

T

)

for 5 cycles

70 % U

T

(30 % dip in U

T

)

for 25 cycles

<5 % U

T

(>95 % dip in U

T

)

for 5 sec

Mains power quality should be that of a

typical commercial or hospital environ-

ment. If the user of the “Mini Daylight”

requires continued operation during

power mains interruptions, it is recom-

mended that the “Mini Daylight” be

powered from an uninterruptible power

supply or a battery.

Power frequency

(50/60 Hz)

magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields

should be at levels characteristic of a

typical location in a typical commercial

or hospital environment.

NOTE U

T

is the a.c. mains voltage prior to application of the test level.