Guidance and manufacturer’s declaration – electromagnetic immunity

The “Mini Daylight” is intended for use in the electromagnetic environment specified below. The customer or

the user of the “Mini Daylight” should ensure that it is used in such an environment.

Immunity test

IEC 60601

test level

Compliance

level

Electromagnetic environment –

guidance

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 V

rms

150 kHz to 80 MHz

3 V/m

80 MHz to 2,5 GHz

3 V

3 V/m

Portable and mobile RF communications

equipment should be used no closer to any

part of the “Mini Daylight”, including cables,

than the recommended separation distance

calculated from the equation applicable to

the frequency of the transmitter.

Recommended separation distance

d=1.2 P

d=1.2 P 80MHz to 800MHz

d=2.3 P 800MHz to 2.5 GHz

where P is the maximum output power rating

of the transmitter in watts (W) according

to the transmitter manufacturer and d is

the recommended separation distance in

metres (m).

Field strengths from fixed RF transmitters,

as determined by an electromagnetic site

survey,

a

should be less than the compliance

level in each frequency range.

b

Interference may occur in the vicinity of

equipment marked with the following

symbol:

 

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorpti-

on and reflection from structures, objects and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land

mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically

with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic

site survey should be considered. If the measured field strength in the location in which the “Mini Daylight” is

used exceeds the applicable RF compliance level above, the Medical Daylight should be observed to verify

normal operation. If abnormal performance is observed, additional measures may be necessary, such as reori-

enting or relocating the “Mini Daylight”.

b

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.