3 fluid information, 4 auto zeroing – Bronkhorst LIQUI-FLOW mini (till 01-07-2013) User Manual

Page 16

Advertising
background image

9.17.065

page 16

Customer model

Type Access

Range

FlowDDE

FLOW-BUS

Unsigned char[16] RW  -

93

113/4


Digital instrument manufacturing configuration information string. This string can be used by Bronkhorst High-Tech to
add extra information to the model number information.

4.1.3 Fluid information

The following parameters give information about the selected fluid range

.

Fluid name

Type Access

Range

FlowDDE

FLOW-BUS

Unsigned char[10]

RW  -

25

1/17


Fluid name consists of the name of the fluid. Up to 10 characters are available for storage of this name.

Fluid unit

Type Access

Range

FlowDDE

FLOW-BUS

Unsigned char[7]

R

-

129

1/31


The Fluid unit can be read by parameter ‘capacity unit’. This parameter contains the unit in maximal 7 characters.

Fluid Capacity (@100%)

Type Access Range

FlowDDE FLOW-BUS

Float R

±1E-10…

±1E+10 21

1/13


Capacity is the maximum value (span) at 100% for direct reading in sensor base units.

Fluid Capacity (@0%)

Type Access Range

FlowDDE FLOW-BUS

Float R

±1E-10…

±1E+10 183

33/22


This is the capacity zero point (offset) for direct reading in sensor base units.

4.1.4 Auto zeroing

To start the auto zero-procedure two parameters should be written:

Control Mode

Type Access

Range

FlowDDE

FLOW-BUS

Unsigned int

RW

0…18

12

1/4


Check section 3.5.2 for available control modes.

Calibration Mode

Type Access

Range

FlowDDE

FLOW-BUS

Unsigned int

RW  9

58

115/1



Value

Mode

Instrument action

0 IDLE

Idle

9 AUTO_ZERO

Auto-zeroing

255 ERROR

Idle