6 removal and return instructions – Bronkhorst LIQUI-FLOW mini (till 01-07-2013) User Manual

Page 21

Advertising
background image

9.17.065

page 21

6 Removal and return instructions

Instrument handlings:

Purge fluid lines

Remove instrument from line

Insert the instrument into a plastic bag and seal the bag

Place the bag in an appropriate shipping container


Add documentation:

Reason of return

Failure symptoms

Contaminated condition

Declaration on Contamination form: 9.17.032


When returning material, always describe the problem and if possible the work to be done, in a covering letter.

It is absolutely required to notify the factory if toxic or dangerous fluids have been metered with the instrument!
This to enable the factory to take sufficient precautionary measures to safeguard the staff in their repair department.
Take proper care of packing, if possible use the original packing box; seal instrument in plastic, etc.

Contaminated instruments must be dispatched with a completely filled in 'declaration on contamination form'.
Contaminated instruments without this declaration will not be accepted.

Note:
If the instruments have been used with toxic or dangerous fluids the customer should pre-clean the instrument.

Important:
Clearly note, on top of the package, the customer clearance number of Bronkhorst High-Tech B.V., namely:

NL801989978B01


If applicable, contact your distributor for local arrangements.

The declaration on contamination form is available at the Bronkhorst download site:
http://www.bronkhorst.com/en/downloads/safety_information_for_returns.pdf