5 troubleshooting and service, 1 general, 2 led indication – Bronkhorst LIQUI-FLOW mini (till 01-07-2013) User Manual

Page 19: 3 troubleshooting summary general

Advertising
background image

9.17.065

page 19

5 Troubleshooting and service

5.1 General

For a correct analysis of the proper operation of a flow meter it is recommended to remove the unit from the process
line and check it without applying fluid supply pressure. In case the unit is dirty or clogged, this can be ascertained
immediately by loosening the fittings and inspecting visually.

Energizing or de-energizing of the instrument indicates whether there is an electronic failure. After that, fluid pressure
is to be applied in order to check behaviour. If there should be suspicion of leakage, do not disassemble the sensor for
inspection but contact your local distributor for service or repairs.

5.2 LED indication

The red LED on the instrument gives error or warning information.

z

z

z

Red LED

Time

Indication

Off Continuously

No

error

On

Continuously

Critical error message
A serious error occurred in the instrument
Instrument needs service before further use

5.3 Troubleshooting summary general

Symptom Possible

cause

Action

No output signal

No power supply

Check power supply and hook-up

Check cable connection and hook-up.

Check status of LED’s
(see manual 9.17.023)

Cable damaged or hooked-up
incorrectly

Check and compare signals at both ends
of cable.

PC-board damaged due to long lasting
shortage and/or high-voltage peaks

Return to factory

No or too low inlet pressure

Increase inlet pressure

Open shut-off at inlet and outlet

Supply pressure too high, or differential
pressure across meter too high

Reduce supply pressure

Sensor / capillary failure

Return to factory

Maximum output signal

Sensor / capillary failure

Return to factory

Output signal much lower than
setpoint signal or desired flow

Incorrect type of liquid or too low inlet
pressure

Test instrument on conditions for which
it was designed

Oscillation / Signal noise

Trapped gas bubbles in the system

Purge the system and make sure no
leaks or dead volumes are present

Pressure regulator of supply pressure is
oscillating or wrong sized

Replace pressure regulator

Small flow indication when
flow is definetly zero

Increased zero reading without flow
caused by raised zero-point

Perform an auto-zero action

No digital communication

Occupied or wrong bus address

Change address with software

For more information check the instruction manual for digital communication 9.17.023 at
http://www.bronkhorst.com/en/downloads/instruction_manuals/

Advertising