Electro-Chemical Devices (ECD) HYDRA Ammonium User Manual

Page 19

Advertising
background image

HYDRA NH

4

+

-N

Page 11

Display variables; which Output to display and which Time interval is used are configured in the Channel

1 Set-Up Menu. See Section 3.2.6, below, for more information on the Set-Up Menu.

3.2.5 Buffer Menu

The Buffer Menus are Calibration Menus. To access the Buffer Menu the cursor must be flashing in front

of the “Ch X Buffer” line. Simultaneously press both of the Horizontal CALIBRATE keys to enter the

Calibration screens. The calibration is structured for a two point calibration.

The first calibration “1 NH4-N” is the “zero point” calibration. This sets a base millivolt value to a specific

ion concentration. The concentration is adjustable to suit the measurement range, see procedure

below. When adjusting a value first move the cursor to the largest integer to be adjusted, perform the

adjustment and then move to the next lower value. When a calibration is performed the Cal line will

display the millivolt value associated with calibration standard solution. The Cal value can be reset to the

Factory default setting by “entering” the Cal line.

The second calibration screen is the “slope, mV/decade” calibration. This calibration should use a

standard solution that is at least 10X the concentration of the solution used in the first calibration. Scroll

down to the “2 NH4-N” line and enter the value of the standard solution and the Cal line will display the

new slope, mV/decade. Follow the same procedure for calibration of Channels 2 & 3.

Screen Displayed

Button Pressed

Ch1 Buffer

CALIBRATE ► (enter the calibration menu by pressing

both ◄► simultaneously)

1 NH4-N 4.00 ppm

Cal .0 mV

CALIBRATE ►(enter line, zero pt. Cal 1)

Sensor should be in the Cal 1 solution.

1 NH4-N 4.00 ppm

Cal 393.5 mV

To set Calibration Sol. Value use the CALIBRATE ◄ or ► to

move cursor. Use the ▲or▼ to adjust the integer value.

1 NH4-N 5.00 ppm

Cal 393.5 mV

MENU SELECT ▼ (accept Calibration)

1 NH4-N 5.00 ppm

Cal 393.5 mV

MENU SELECT ▼ (move to Cal line)

1 NH4-N 5.00 ppm

Cal 393.5 mV

MENU SELECT ▼(move to Cal 2, Slope)

Remove sensor, rinse and place in Cal 2 solution

2 NH4-N 00.0 ppm

Cal 29.6 mV/dec

CALIBRATE ► (enter screen)

2 NH4-N 00.0 ppm

Cal 29.6 mV/dec

To set Calibration Sol. Value use the CALIBRATE ◄ or ► to

move cursor. Use the ▲or▼ to adjust the integer value.

2 NH4-N 50.0 ppm

Cal 29.6 mV/dec

MENU SELECT ▼ (accept Calibration)

2 NH4-N 50.0 ppm

Cal 29.2 mV/dec

CALIBRATE ▼ (return to Home Screen)

3.2.6 Set-Up Menu

The Setup menu allows the various parameters to be configured for the user’s requirements. Each

Channel has a Set-Up Menu that includes the parameters associated with that specific Channel. The

Advertising
This manual is related to the following products: