Electro-Chemical Devices (ECD) HYDRA Ammonium User Manual

Page 33

Advertising
background image

HYDRA NH

4

+

-N

Page 25

4.2.4 Two point Calibration

The second point of a Two Point Calibration sets the slope of the sensor, the mV per decade. The slope is

calculated by comparing the millivolt values and ppm values in the “1NH4-N 1.00 ppm” line to the

values in the “2NH4-N 100 ppm” line of the Buffer Menu. The concentration of the solution used for

the second point should be at least 10 times higher than the value used in the first point of the

calibration. The recommended calibration standards for both the ammonium ion and potassium ion

calibrations are 1 ppm and 100 ppm. Perform a two point calibration whenever an electrode is replaced.

After completing the single point calibration, see section 4.2.3 above, rinse the sensor with distilled

water and gently dab it dry with a paper towel or soft tissue. Carefully cleaning the sensor prevents

errors in the calibration due to carryover from the first solution. Place the sensor in the second solution,

either the pH 4.01 buffer or one of the 100 ppm solutions and wait for the reading to stabilize before

accepting the value. See section 3.2.5 above for further instruction.

4.2.5 Potassium Ion Compensation

Potassium ions have a positive interference on the Ammonium Ion Electrode, i.e. the ammonium

reading is higher than the actual value. The Potassium Ion Compensation, K+ comp, adjusts the

measured potassium ion concentration using the interference ratio value, the Ks value, and subtracts

the resulting amount from the Ammonium Ion Signal. The Ks value in the NH4-N Set Up screen allows

the interference ratio to be adjusted as needed. The default setting is Ks = .100 which is 10 potassium

ions produce the same signal as 1 ammonium ion, Ks = 0.04 would change the correction to 25:1. The

0.100 correction factor is good for most solutions with low levels of potassium, less than 20 ppm.

Each Potassium Ion Electrode should have similar characteristics but adjustments are necessary when

the potassium electrode is replaced or when high levels of potassium are present in the solution. If the

ammonium ion concentration is less than expected when compared to a grab sample measurement

then adjust the Ks value to a smaller value. Adjusting from 0.100 > 0.082 for example would increase the

potassium to ammonium ion ratio from 10:1 to 12:1 thereby slightly increasing the displayed

ammonium ion concentration by subtracting less in the compensation.

Advertising
This manual is related to the following products: