Electro-Chemical Devices (ECD) HYDRA Ammonium User Manual

Page 32

Advertising
background image

Page 24

HYDRA NH

4

+

-N

4.2.2 Temperature Calibration

Place the HYDRA sensor in a beaker of water so that the bottom 3 inches (7.6 cm) of the body is

immersed. Allow the sensor to equilibrate for 10 minutes. Verify the displayed temperature agrees with

the thermometer. If not, then adjust the temperature in the Configure/Trim > °C/°F & temp cal > 1 Trim

°C menu to agree with the thermometer. This calibration was performed in the factory calibration but it

is advisable to perform a temperature check every 2-3 months.

4.2.3 Single point Calibration

Single Point Calibrations are used to standardize the sensor’s reading to a known value typically to

eliminate errors caused by drift or large changes in the makeup of the measured solution. A Single Point

Calibration adjusts the base potential of the sensor by shifting the response curve to a higher or lower

level but it does not affect the slope (mV/ppm) of the curve.

Single point calibration is available for each of the measured parameters in the appropriate Buffer

Menu, Channel 1 Buffer for the NH

4

+

-N, Channel 2 Buffer for the K

+

and Channel 3 Buffer for pH. (See the

Single Point Calibration Screen table below)

Immerse the front end of the sensor into a container of the calibration solution, enter the value of the

calibration solution or the value derived from a grab sample into the “1 (ion) xxx ppm” line of the

Buffer menu as described in Section 3.2.5 above. Accept the reading when stable by pressing either of

the MENU SELECT keys. The lower line in the menu, the Cal line will show the new mV value associated

with the entry.

***Important Note for Ammonium Calibrations*** When using the 1 ppm and 100 ppm NH

4

+

-N

calibration solutions to calibrate the HYDRA C22 analyzer the potassium ion compensation, the “K+

comp” line in the Channel 1 Setup menu should be turned on, it is on by default. These calibration

solutions contain no potassium ions but the ammonium ions interfere with the potassium ion electrode

and the potassium ion electrode registers a K

+

reading, this value must be accounted for in the NH

4

+

-N

calibration.

If the Single Point Calibration is the first point in a Two Point Calibration it is customary to use the

solution with the lowest value for the single point calibration. There are no restrictions on the value

entered in the “1 (ion) xxx ppm” line of the Buffer menu; it can be higher or lower than the value in line

2 of the menu. For the pH calibration the zero point is set with pH 7.0 calibration buffer and the slope is

set with the pH 4.01 calibration buffer.

Single Point Calibration Screens

Channel

Screen

Recommended Solution

Ch1 Buffer (NH4-N)

1NH4-N 1.000 ppm
Cal 320.0 mV

1.0 ppm NH4-N (Part# 2010445)
or Grab Sample value

Ch2 Buffer (K+)

1K+ 1.000 ppm
Cal 165 mV

1.0 ppm K (Part# 2010443)
or Grab Sample value

Ch3 Buffer (pH)

1 pH 7.00 pH
Cal 3.1 mV

pH 7.0 Buffer (part# 2010101)
or Grab Sample value

Advertising
This manual is related to the following products: