Electro-Chemical Devices (ECD) HYDRA Ammonium User Manual

Page 21

Advertising
background image

HYDRA NH

4

+

-N

Page 13

Timer 1: periodic
←Clock

CALIBRATE ►

Per 0d00 :00 :00 Cleaning Timer
Now 0 00 :00 :00 (Per sets the time
On 0 00 :00 :00 between cleaning
Off 0 00 :00 :00 cycles)

CALIBRATE ►

(Set the Period to 15 minutes then every ¼ hour the Now
Clock will reset to 0:00:00 and trigger the On Timer which is
set at 0:00:00, starting the cleaning cycle.)

Per 0d00 :15 :00
Now 0 00 :00 :00
On 0 00 :00 :00
Off 0 00 :00 :00

To adjust the value use the CALIBRATE ◄ or ► to move
cursor. Use the ▲or▼ to adjust the integer value, then
MENU SELECT ▼ (save the Value)
MENU SELECT ▼, MENU SELECT ▼, MENU SELECT ▼

Per 0d00 :15 :00
Now 0 00 :00 :00 (Set the Off time
On 0 00 :00 :00 at 30 seconds)
Off 0 00 :00 :30

CALIBRATE ► (enter the Off line, the duration of the

cleaning cycle) To adjust the value use the CALIBRATE ◄ or

to move cursor. Use the ▲or▼ to adjust the integer

value, then MENU SELECT ▼ (save the Value)

Timer 2: oneshot↑
←timer 1←clock

CALIBRATE ►

4-20 mA hold time after cleaning cycle, (CT + 1 minute)

Now 0 00 :00 :00 Manual Mode Timer
On 0 00 :00 :00 (Hold Function)
Off 0 00 :01 :30 (Set at 1 min 30 sec)

MENU SELECT ▼, MENU SELECT ▼, ◄ CALIBRATE ►
To adjust the value use the CALIBRATE ◄ or ► to move
cursor. Use the ▲or▼ to adjust the integer value.

Ks .120
K+ comp Off

MENU SELECT ▼ NH

4

electrode selectivity ratio for K

+

NH

4

reading – 0.120 (K

+

reading) = NH

4

concentration

Ks .120
K+ comp Off

MENU SELECT ▼ K

+

compensation should be Off for

Calibration and On for Measurement

TC .333 %/°C
Isopot 20.0 mV
Equilib 9.24 pH
Dissoc 1.000 Off

MENU SELECT ▼

Temperature Compensation (no adjustments needed)

TC .333 %/°C
Isopot .0 mV ** Mandatory **
Equilib 9.24 pH ** Input **
Dissoc 1.000 Off

CALIBRATE ► (Enter Iso-Potential Entry line)


**Iso-Potential value supplied with each electrode**

TC .333 %/°C
Isopot 20.0 mV
Equilib 9.24 pH
Dissoc 1.000 Off

To adjust the value use the CALIBRATE ◄ or ► to move
cursor. Use the ▲or▼ to adjust the integer value.

TC .333 %/°C
Isopot 187.0 mV
Equilib 9.24 pH
Dissoc 1.000 Off

MENU SELECT ▼ (accept the Value)
MENU SELECT ▼

Equilibrium value, No Input Needed

(Equilib= pH value equal to the pKa, the acid dissociation
constant, where [NH

4

= NH

3

+ H

+

])

TC .333 %/°C
Isopot 187.0mV
Equilib 9.24 pH
Dissoc 1.000 On

MENU SELECT ▼
Dissoc = the pH correction factor for the [NH

4

↔ NH

3

+ H

+

]

dissociation , (the Measured NH

4

)/Dissoc = Total NH

4

-N

(should be on at all times)

Noise Filter 5