2.3 Analysis Cycle

A typical Analysis Program in the Model CA-6 would have the following structure: Rinse the colorimetric
reaction cell and take a sample, add one or more reagents like a buffer or masking agent and then make
the first measurement, the reference measurement. The reference measurement eliminates interfering
factors such as sample color and turbidity, miscellaneous color from the reagents and refractive index
variations.
After the reference reading, the color producing reagents are added. The sample is mixed and allowed
time to complete the color forming reactions before taking the second measurement, the Reading
measurement
. The reference and reading values are used to calculate the concentration using the
calibration factor. The reaction cell is drained and rinsed several times before starting the next
measurement.

Fig. 2-3 Flow diagram

2.3.1 Typical Run Sequence:

Conditioning, rinsing and sampling

Drain, rinse and sample functions

First the cuvette is drained and rinsed (these steps can also be

programmed at the end of the run). The hydraulic lines and the

colorimetric cell are rinsed prior to taking the actual sample. Then

the sample is taken.

Addition of reagent(s)

Add reag function

Depending on the method one or more reagents are added before

the reference reading.

Mixing and wait

The mixing pump is activated and the liquid is pumped from the

lower part to the upper part of the colorimetric cell. The waiting

9