Electro-Chemical Devices (ECD) CA6 Hardness Analyzer User Manual

Page 23

Advertising
background image

Sample #1

Sample #1 actuates the peristaltic pump with the bottom pinch valve closed

and the colorimetric cell fills with sample #1. The sample time is set in

seconds.

SAMPLE #1 = left side of the selection valve

Sample #2
Sample #2 actuates the peristaltic pump with the bottom pinch valve closed

and the colorimetric cell fills with sample #2. The sample time is set in

seconds.

SAMPLE #2 = right side of the selection valve

Dilution options

Loop On
Loop On allows the sample liquid (sample line #1) that is to be diluted to

flow through the Sample Loop. The Loop On time is set in seconds.
Loop off
Loop Off traps the sample inside the loop tubing while the sample fluid is

flushed from the lines with the dilution water. The Loop off time is set in

seconds.
Loop On
Loop On for the dilution fluid, DI water (sample line #2), dilutes the sample

in the loop tube and fills the Colorimetric Cell.

12