Fieldpiece SDP2 - Dual In-Duct Psychrometer User Manual

Fieldpiece, Dual in-duct psychrometer, Quick start

Advertising
background image

01

03

06

02

05

04

07

09

12

08

11

10

Quick Start

1. Power on your SDP2 by holding the

button for 1 second.

2. Select mode with the TEET/ΔT

button: Target Evaporator Exit

Temperature, Target Temp Split, or

Normal (real-time return and supply.)

3. Press the PARAM button to cycle

between dry bulb, relative humidity,

wet bulb, dew point, and enthalpy.

4. Insert sensing probes into duct or

plenum for in-duct measurements.

5. Hold SYNC to pair and send all

measurements to Fieldpiece model

HG3 or SMAN4 Analyzers.

Certifications

C-Tick (N22675)

CE

WEEE

FCC


RoHS Compliant

Description

The two probes each simultane-

ously measure temperature and

RH%. One goes before the evapo-

rator (RETURN) and one goes after

the evaporator (SUPPLY). These four

measurements can be displayed or

used in calculations to display the

actual temperature split (delta T), the

target temperature split, the actual

exit evaporator temperature, or the

target exit evaporator temperature

(TEET), plus the difference between

the actual and target. Enthalpy (BTU/

LBM) and Dew Point can also be dis-

played.

The TEET takes into consideration

the latent heat used to condense

water from the air, while a simple 20°

temperature split ignores latent heat.

TEET, developed by Fieldpiece, is an

intuitive test allowing you to aim for

a target value rather than a target

difference between two values.

The 38"(96cm) telescoping probe

with laser etched ruling and flattened

edges allows you to locate proper

measurement points within a duct

and ensures that your probe is prop-

erly aligned. The dual display, bright

backlight, and rugged rubber boot

with probe clips make sure the SDP2

is ready for any job.

Maintenance

Clean the exterior with a dry cloth.

Do not use liquid.

Battery Replacement

Turn your SDP2 off, remove rub-

ber boot, unscrew battery cover, and

replace 9V battery.

Fieldpiece

Dual In-Duct

Psychrometer

OPERATOR'S

MANUAL

Model SDP2

Display

Auto Power Off Enabled

Battery Life

Wireless Signal Strength

Relative Humidity

Dew Point

Wet Bulb

Dry Bulb

Enthalpy

Target Evaporator Exit

Temperature

Target Dry Bulb Temp Split

Temperature

Maximum Display

Minimum Display

Hold Display

Δ Display (Return-Supply)

Top Display Shows Return

Bottom Display Shows Supply

or the difference between actual and target

values for TEET and ΔT.

Controls

Hold 1 second to toggle power on/off.

Press to toggle backlight.

Cycle through dry bulb, wet bulb, dew

point, relative humidity, and Enthalpy.

Hold for 1 second to search for a model

HG3 or SMAN4 and send real-time

measurements.

Cycles through Hold, Difference (Δ),

Maximum, Minimum, and Real-time.

Hold for 1 second to exit and clear stored

values. Hold while turning ON to toggle

Auto-Off (APO).

Press to cycle between Target

Evaporator Exit Temp,

Target Delta T and Normal

modes.

Press both to toggle between

Fahrenheit or Celsius and BTU/

LBM or KJ/KG.

BACKLIGHT NOTE: The timer is automatically reset

for 3 min when any button is pressed. A quick press of

toggles backlight.

How to Use

Normal Mode

Measure real-time dry bulb, rela-

tive humidity, wet bulb, dew point, or

BTU/LBM for each probe.

1. Remove vinyl protective sleeves if covering sensors.

2. Press the PARAM button to cycle dry bulb

temperature, relative humidity, wet bulb, dew

point, or BTU/LBM.

3. The top display shows Return probe and the

bottom display shows Supply probe. The probes are

interchangeable.

TEET (Target Evaporator

Exit Temperature) Mode

Target Evaporator Exit Temperature,

is similar to temperature drop or ΔT.

TEET is better than a simple temp-split

because it uses a wet and dry bulb

temperature on the evaporator coil

to determine the coil load and then

calculates a corresponding correct

exit dry bulb temperature.

Insert the Return probe to find the

TEET. Insert the Supply probe to moni-

tor how close the target temperature

is to the actual temperature.

1. Insert the RETURN probe into the return plenum.

See Figure 1 for recommended return plenum

placement of probe. Drill or punch a 3/8" hole into

the return plenum if needed.

2. Press the TEET/ΔT button until TEET is displayed on

the top line.

3. Insert the SUPPLY probe into the supply plenum

to see how close the target temperature is to the

actual temperature. See Figure 1 for recommended

supply plenum placement of probe. Drill or punch a

3/8" hole into the supply plenum if needed.

4. (Actual evaporator temp - target evaporator

temp) will be shown on the bottom "SUPPLY"

line. A negative number means the actual exit

temperature and airflow are both too low. You

want to be within ± 3°F (± 1.7°C) of the target for

good evaporator performance.

5. Seal any holes before leaving the jobsite.

Target ΔT (Target Temp-

Split) Mode

Temperature split is simple. It is the

return temperature minus the supply

temperature. Target ΔT uses RETURN

wet bulb and dry bulb to measure

heat load and automatically calculate

what the ΔT should be.

Insert the RETURN probe to find

the Target ΔT. Switch over to Normal

mode, and use both probes to then

see what your dry bulb Δ actually is.

1. Insert the RETURN probe into the return plenum.

See Figure 1 for recommended return plenum

placement of probe. Drill or punch

!

a 3/8" hole

into the return plenum if needed.

2. Press the TEET/ΔT button until TARGET ΔT is

displayed on the top line.

3. Switch to Normal Mode to find the actual ΔT so you

can compare the two values.

4. Insert the SUPPLY probe into the supply plenum.

See Figure 1 for recommended supply plenum

placement of probe. Drill or punch

!

a 3/8" hole

into the supply plenum if needed.

5. Press PARAM until DB is shown. Press H/Δ/M/M

until Δ is shown.

6. Seal any holes before leaving the jobsite.

RCONE1 Probe Lock

Figure 2. Screw a threaded RCONE1

into the duct or plenum wall to secure

a sensing probe. Use the RCONE1s

and the magnet on the SDP2 for

hands-free in-duct testing.

Wireless Sync

The SDP2 can wirelessly send all of

its measurements to a wireless Field-

piece model that accepts them. As of

this printing, models HG3 and SMAN4

both can receive measurements.

1. Enter Target Superheat mode.

2. Use arrow keys to select IDWB. (IDWB will blink)

3. Press SYNC on SMAN4 for 1 second.

4. Press SYNC on SDP2 for 1 second to connect and

sync real-time indoor wet bulb.

5. Use a temperature accessory head and ET2W wire-

less transmitter to sync ODDB and view real-time

target superheat while you charge outside.

1. Select test.

2. Use arrow keys to select a measurement line.

3. Press SYNC on HG3 for 1 second.

4. Press SYNC on SDP2 for 1 second to connect and

sync ALL lines of the test that it can fill.

5. Values will be held if OUTPUT is pressed. Going back

to INPUT will show the held values. Press SYNC on

the HG3 to return to real-time values.

80

80

80

80

80

80

80

80

80

80

80

80

80

80

80

80

80

80

80

80

80

80

80

80

80

80

80

80

80

80

80

80

80

80

80

80

80

80

Figure 1

Figure 2

ARH5

In-Duct

Psychrometer

Head

ON

AUTO OFF

1%RH/mVDC

1°F/mVDC

1°C/mVDC

LO BATT

DEW POINT

%RH

WET BULB

TEMP

°F

°C

ON

OFF

ET2W

LO BATT

SEND

RECEIVE

ON

AC

DC

Wireless

Transmitter

SYNC

Clear

Input

Output

Recall

Enter

Save

Sync

SETUP

Target

Evaporator

Logger

Light

Exit Temp

CFM

Display

Superheat

Subcooling

Combustion

HVAC Guide

CheckMe!

Fieldpiece

HG3

Data

Service

E

Exit T

Ex

Ex

E

e

Target

arget

arget

rget

rget

get

Targe

or

ator

ato

ator

ator

ator

Ta

Evaporat

t Temp

INPUT FORM

SH Table: Standard

Refrigerant: R-22

OD Dry Bulb:

85.4°F

Return WB:

63.0°F

SL Pressure:

52.4psig

SL Temp:

45.4°F

CustomerID:

__JONES123

Use the RCONE1 for

hands-free in-duct

measuremesnts.

PARAM

80

80

H/Δ/M/M

TEET/ΔT

SYNC

PARAM

SYNC

+

Clea

r

Input

Output

Recall

Enter

Save

Sync

SETUP

Target

Evaporator

Logger

Light

Exit Temp

CFM

Display

Superheat

Subcooling

Combustion

HVAC Guide

CheckMe!

Fieldpiece

HG3

Data

Service

Display

INPUT FORM

SET 01

Set ID:Indoor Temp

Measurement:

72.8

SET 02

SET ID:Comp Amps

Measurement:

23.6

CustomerID

: __JONES123

HVAC Guide® System Analyzer

Model HG3

4-Port Wireless Manifold

Model SMAN4

WARNING

Do not retract the sensor probe by pulling on the

cord. Doing so may sever the cord from the sensors.

Do not punch or drill holes into an evaporator

without making sure there is not critical system parts

behind the hole location.

!

!

Recommended hole
placement for RETURN
measurement.

Recommended hole
placement for SUPPLY
measurement.

Look behind the
panel before drilling
and make sure you
do NOT drill a hole
through any critical
components.

!

!

Advertising