Flowserve 505si Series Digital Positioner User Manual

Page 8

Advertising
background image

8

®

User Instructions Logix 505si - LGENIM0505-00 03/09

b.

Split range - Split range selection is used to limit the

action control region to either 4 - 12 mA or 12 - 20 mA.

Thisallows two valves to be controled from the same

4 - 20 mA control line.

Full: The 505si uses the full 4 - 20 mA range to control

the valve.

Split: The 505si uses either 4 - 12 or 12 - 20 mA to

control the valve.

c.

4 - 12 (12 - 4) or 12 - 20 (20 - 12) is used to select the

range used for control. This switch is only active when

the full/split switch is in the split position.

4 - 12 (12 - 4) Full valve travel in the 4 - 12 mA range

for ATO (12 - 4 mA range for ATC)

12 - 20 (20 - 12) Full valve travel in the 12 - 20 mA

range for ATO (20 - 12 mA range for ATC)

d.

gain Switch – This switch adjusts the position control
algorithm of the positioner for standard or low gain.

Standard: Placing the switch to the left opmizes

the response for low friction, high performance control

valves. This setting provides for optimum response

times when used with most low friction control valves.

Low: Placing the switch to the right reduces the gain. It

also optimizes the response for valves and actuators with

high friction levels. This setting slightly slows the

response and will normally stop limit cycling that can

occur on high friction valves.

WARning: During the Quick-Cal operation the valve

may stroke unexpectedly. Notify proper personnel

that the valve will stroke, and make sure the valve is

properly isolated.

8.3

QUiCK-CAL Operation

The QUICK-CAL button is used to locally initiate a

calibration of the positioner. Pressing and holding the

QUICK-CAL button for approximately 3 seconds will ini-

tiate the calibration. The settings of all the configuration

switches are read and the operation of the positioner

adjusted accordingly. A QUICK-CAL can be aborted at

any time by briefly pressing the QUICK-CAL button and

the previous settings will be retained. While the calibra-

tion is in progress you will notice a series of different

lights flashing indicating the calibration progress. When

the lights return to a sequence that starts with a green

light the calibration is complete. (see the section 9 for

an explanation of the various light sequences)

8.4

Factory reset

Hold Quick cal button while applying power and all of

the internal variables including calibration will be reset

to factory defaults. The positioner must be re-calibrated

after a factory reset.

Figure 4: Logix 505si Local interface

Configuration

Switches

4-20 mA Input

LED Status

Lights

Quick-Cal

Switch

STOP!

Advertising