Nuclear Associates Radiographic and Mammographic Focal Spot Measurement Products
Operators Manual

2-6

[

]

[

]

T = tube assembly tilt angle (Provided by the x-ray system manufacturer.)

For example, for a 0.80 mm measured focal spot length (F), an anode angle (A) of 20°, and a tube
assembly tilt angle of 8°, one obtains

Sin (20°)

L = 0.80 mm

Sin (8° +

20°)

0.34

L = 0.80 mm

0.47

= 0.58

mm

14. Compare the dimensions to the focal spot size tolerance limits in Table 2-2. The manufacturer may

not accept screen-film slit image measurements for acceptance testing purposes. In this case, it
may be necessary to repeat the procedure using direct exposure x-ray film.

Table 2-2.

Focal Spot Size Tolerance Limits

(For Slit Camera Method)

Maximum Focal Spot

Dimensions in mm (F

eff

)

Nominal Focal Spot

Designation (F)

Width Length

0.05 0.075

0.075

0.1 0.15

0.15

0.15 0.23

0.23

0.2 0.30

0.30

0.25 0.40

0.40

0.3 0.45

0.65

0.4 0.60

0.85

0.5 0.75

1.1

0.6 0.90

1.3

0.7 1.1

1.5

0.8 1.2

1.6

0.9 1.3

1.8

1.0 1.4

2.0

1.1 1.5

2.2

1.2 1.7

2.4

1.3 1.8

2.6

1.4 1.9

2.8

1.5 2.0

3.0

1.6 2.1

3.1

1.7 2.2

3.2

1.8 2.3

3.3

1.9 2.4

3.5

2.0 2.6

3.7