Nuclear Associates Radiographic and Mammographic Focal Spot Measurement Products
Operators Manual

2-12

5. For the Star Test Pattern, the formula and data

presented yield the equivalent homogeneous
focal spot size, not the actual size.

2.4 Procedure - Half-Value Layer Measurements

1. Place the focal spot test stand on the tabletop. Adjust the leveling screws so that the base is raised

1" (2.5 cm) above tabletop.

2. Place a sheet of lead under the base.

3. Adjust the stand so that the top is located at one-half of source-to-tabletop distance, e.g., for a 40"

source-to-table-top distance adjust the stand to 20".

4. Place both the large and small adaptors in the top of the stand.

5. Collimate the x-ray beam to the top aperture of stand.

6. Place the ionization chamber on the base just above the aperture (Figure 2-11).

7. Select the technique factors that will produce about 300 mR at 80 kVp.

8. Make an exposure without added aluminum and record the result.

9. Add 2 mm of type 1100 aluminum.

9. Make an exposure and record the result.

10. Repeat steps 9 and 10, adding an additional 1 mm of aluminum each time until the exposure

reading is approximately 100 mR.

12. For data analysis, potential problems, and acceptance limits, See Reference 2 on page 2-12.