Operation

X-Ray Tube Focal Spot Measurements

2

2-1

Section 2

Operation

2.1 X-Ray Tube Focal Spot Measurements

2.1.1 Purpose

To assure that the tube and x-ray tube focal spot size is within acceptable limits.

2.1.2 Equipment Needed/Recommended for X-Ray Tube Focal Spot

Measurements

1. Slit assembly, Pinhole, or Star Pattern.

2. A Focal Spot Test Stand (See Figure 2-1), with small and large adapter rings, and fluorescent

screen.

3. Focal Spot Test Stand Alignment Device (Figure 2-2).

4. 8" X 10" (20 x 25 cm) screen-film extremity or mammography cassette and film, or direct-exposure

x-ray film.

5. Clear plastic metric ruler.

6. Spirit

level.

7. 6X magnifier with graticule scale in 0.1 mm divisions.





















Figure 2-1. Multipurpose Focal Spot Test Stand (Model 07-622)