Fluke Biomedical 35035 User Manual

Page 14

Advertising
background image

35035
Operators Manual

4-2

time specified and verify that the Model 35035 display reads within the limits listed. Reversing the polarity
of the current source signal should reverse the polarity indicated on the Model 35035 display without
significantly altering the numeric value for either mA or mAs. On mAs, the reading generated by a 50 mA,
5 s pulse should be effectively nulled out by an equal pulse of reverse polarity. (Note that except when
performing the mAs tracking test, the RESET button must be depressed after each measurement
reading.)

4.4.1 MA Performance Tests

35035 Range Setting

Current Source Output

35035 Display Reading

20

5 mA

4.95 - 5.05

20

15 mA

14.85 - 15.15

200

50 mA

49.5 - 50.5

200

150 mA

148.5 - 151.5

2000

500 mA

495 - 505

2000

1500 mA

1485 - 1515

4.4.2 MAS Performance Tests

35035 Range Setting

Current Source Output

Current Source

Dwell Time

35035 Display

Reading

20 15

mA

1

s

0

200

50 mA

1 s

49.5 - 50.5

200

500 mA

.1 s

49.5 - 50.5

200

1500 mA

.1 s

148.5 - 151.5

2000

50 mA

5 s

248 - 253

2000

500 mA

1 s

495 - 505

2000 1500

mA 1

s

1485

-

1515




The following test will verify the mAs tracking
performance. Do not depress the RESET button
between applied current pulses.

2000

50 mA

5 s

DO NOT RESET!

2000 -50

mA

5

s

LESS THAN ± 005